1. Probe Model: Mindray C5-2E for DC-8 2. Fault Detection Results: Upper case of crystal broken Housing broken and color changed Color of cable outer skin changed Lens damaged 3. Repair Parts Crystal Crystal side strain relief Housing 4. Before Repair & After Repair Please click here to view more ...
1. Probe Model: GE IC5-9-D for LOGIQ E9 2. Fault Detection Results: Outer skin color of the cable changed Skin color of crystal side strain relief changed Lens broken 3. Repair Parts Lens replacement 4. Before Repair & After Repair Please click here to view more of MYPRO repair cases.
1. Probe Model: Samsung Medison CV1-8AD for HS60 2.Fault Detection Results: Outer skin color of the cable changed Dome damaged Oil leakage 3. Repair Parts Dome replacement Repair oil leakage 4. Before Repair & After Repair Please click here to view more of MYPRO repair cases.
1. Probe Model: Siemens 4C1 for S2000 2.Fault Detection Results: Outer skin color of the cable changed Lens damaged 5 elements weak for echo test Shadow in air scan image 3. Repair Parts Crystal replacement 4. Before Repair & After Repair Please click here to view more of MYPRO repair cases.
1. Probe Model: GE C1-5-D for Logiq S8 2.Fault Detection Results: Scratches on scanhead front case Outer skin color of the cable changed Lens damaged 3 elements weak for echo test 3. Repair Parts Lens replacement 4. Before Repair & After Repair Please click here to view more of MYPRO repair cases....
1. Probe Model: Mindray L14-5WU for Resona 7 2.Fault Detection Results: Color of cable outer skin changed Lens damaged 9 elements weak and 5 elements missing for echo test 3. Repair Parts Lens 4. Before Repair & After Repair Please click here to view more of MYPRO repair cases.
1. Probe Model: Toshiba PVM-375AT for SSA-550A 2.Fault Detection Results: Crystal housing craze Outer skin of cable is aging and turning yellow, severely hardening Lens is aging There is dark area and obvious interference in air scan image 3. Repair Parts Crystal Cable Crystal case and housing ...
1. Probe Model: Philips C5-1 for CX50 2.Fault Detection Results: No lens; Color of the outer skin of cable changed; Color of the strain relief for crystal side changed; The capacitance test shows 5 elements weak, 3 elements weak without crystal; 2 elements weak and 2 elements missing for echo test; ...
Yes, ultrasound probes can be repaired. But it depends on the nature of the damage and the type of probe. There are several common issues that may occur with ultrasound probes, such as: 1. Cracked or Broken Housing: If the outer casing or housing is cracked, it can often be replaced. A technician ...