| Brand Name: | MYPRO |
| Model Number: | GE 9L-D Crystal Replacement |
| MOQ: | 1 Piece |
| Price: | Negotiable |
| Payment Terms: | T/T,Western Union,PayPal |
| Supply Ability: | 200 Pieces per Month |
Piezoelectric crystal replacement for the GE 9L-D linear probe is a repair path built for vascular, musculoskeletal, and small parts imaging, where a single dead element shows up immediately as a vertical dropout. The 9L-D array is dense and thin, so a failed crystal row cannot simply be bridged or bypassed. The array has to be opened, the damaged elements removed, and new crystals lapped to the exact thickness that restores the original center frequency. Our workshop performs that operation in house, then re-laminates the matching layer and refits a new lens so the probe returns to clinical duty with a uniform element response instead of a patchwork image.
| Parameter | Specification |
|---|---|
| Compatible Probe Model | GE 9L-D linear (also 8L-RS, 11L-D, 12L-RS on request) |
| Probe Type | Linear array, vascular and small parts |
| Array Elements | 192 elements, crystal replacement in single element or continuous block groups |
| Nominal Frequency | 4.0 to 10.0 MHz, crystal thickness matched to original center frequency |
| Element Pitch | Original pitch reproduced within 0.01 mm tolerance |
| Replacement Crystal | Medical grade PZT-5H equivalent, capacitance sorted in 5 percent bands |
| Matching Layer | Two-stage acoustic matching layer, re-bonded and vacuum degassed |
| Acoustic Lens | New molded lens, RTV bonded, cure controlled at low temperature |
| Sealing | Immersion test 30 minutes at 1 m on every finished unit |
| Verification | Element map, sensitivity spread, wire phantom resolution set |
| Turnaround | 5 to 7 working days for standard faults, 3 days expedited |
| Warranty | 12 months on replaced crystals and lens bond |
The 9L-D is used daily for carotid and peripheral vascular studies, thyroid and breast work, and musculoskeletal injections, so a striped image is a scheduling problem rather than a cosmetic one. Crystal replacement keeps the original connector, cable, and housing in service, which matters for departments that have already validated the 9L-D preset on their GE system and do not want to re-qualify a new probe model. Refurbishment companies use the same service to convert cores with known element faults into sellable inventory, and mobile imaging providers rely on it to keep a small probe pool in rotation without buying new transducers.
When the whole array has failed rather than a localized group, we report that honestly and recommend replacement instead of charging for a rebuild that cannot meet specification.
Finished probes travel in a foam-cradled export carton with the element map, immersion test record, and wire phantom images enclosed. Rework is performed in a humidity-controlled area because moisture ingress during crystal bonding is the main cause of premature element failure in rebuilt linear arrays. Each replacement crystal is checked for capacitance and pulse-echo response before bonding and again after cure, and any element that drifts outside the acceptance band is removed and replaced rather than accepted with a note. Warranty covers crystal and lens bond defects for twelve months, and returned units are re-mapped and re-bonded without additional labour charges.