| Brand Name: | MYPRO |
| Model Number: | PVT-675MVL |
| MOQ: | 1 Piece |
| Price: | Negotiable |
| Payment Terms: | T/T |
| Supply Ability: | 300 probes per month |
An error message on the console rarely names the real fault. MYPRO starts every repair with diagnosis: element activity mapping, cable continuity and impedance measurement, leakage current test and a phantom image baseline. You receive a written report that translates the console error into an actual probe fault before any repair cost is committed.
| Probe Model | Toshiba PVT-675MV 4D convex array (Aplio series) |
|---|---|
| Diagnosis Scope | Element map, cable continuity, impedance, leakage current, acoustic output, phantom imaging |
| Repair Scope | Dead element and crystal array replacement, lens and dome replacement, cable and connector repair, oil filling |
| Deliverable | Written fault report linking console error code to the failed component |
| Turnaround Time | Diagnosis within 24 hours of arrival, repair 5–7 working days |
| Warranty | 12 months on replaced parts and workmanship |
| MOQ | 1 probe |
| Shipping | DHL, FedEx, UPS |
Send the exact console error text, the system model and a photo of the probe label, and we will tell you what the error most likely means before shipping.